奥谱天成(厦门)科技有限公司
Optosky Technology Co., Ltd.

服务热线:
0592-6102588
ATP1100-生化分析光谱仪
|
型号 : ATP1100
产品详情

GENERALDESCRIPTION

Optosky’s ATP1100 is customized for automatic biochemical analyzer. It features crossed asymmetricalCzerny-Turner optical design.

Thedetectable wavelength is determined by the pixel arrangement and quantity ofthe PD array. The light collection can be finished within 1.2 ms, and the data processingand transmission can be finished within 5 ms.

ATP1100is optimized the efficiency in UV spectral region. The integration time can beset through the communication interface RS485. ATP1100 begins to collect light signalon the TTL rising edge. And then analysis and output the AD data via RS485/232.

FEATURES

lMeet with China standard YY/T0654-2008

lFree space Interface with 25 mm focuslength

lWide spectral range: 340800 nm

l10 or 12 detectable wavelength: 340nm, 405 nm, 450 nm, 480 nm, 505 nm, 546 nm, 570 nm, 605 nm, 660 nm, 700 nm, 750nm, 800 nm

lUV enhanced response

lCustomized 10 or 12-pixel PD array

lExcellent stray light eliminateability

lLow noise circuit

lPower: single +12V DC @ 1.0A

lTrigger: TTL rising-edge

lOutput: RS485/232

APPLICATIONS

lBiochemical analyzer



1      OpticalPerformance

Thelinearity performance of ATP1100 was evaluated by testing the 340 nm absorbedsolution with volume concentration from 1/10 to 10/10. The testing resultsshowed in the following figure1 are obtained by three sets of ATP1100spectrometers. The accuracy is better than 5% and 7% whenthe nominal   absorbance of the testingsolution is 1 and 2.5. The testing results show that ATP1100 can meet with thepharmaceutical industry standard of China (YY/T0654-2008).

Figure.1The absorbance results using 3 different sets of ATP1100 vs the theoreticvalues. The results shows that ATP1100 can meet with thepharmaceutical standard of China(YY/T 0654-2008).

Table 1 The testing accuracy of three different sets of ATP1100

Volume   concentration of standard samples

Theoretic   absorbance of standard samples

Measurement   error of   #1 ATP1100

Measurement   error of   #2 ATP1100

Measurement   error of   #3 ATP1100

1/10

0.336

0.310%

-0.945%

-0.527%

2/10

0.672

0.112%

-1.672%

-0.072%

3/10

0.987

-0.703%

1.465%

0.686%

4/10

1.340

-0.485%

-0.598%

1.242%

5/10

1.674

0.578%

-0.505%

1.889%

6/10

2.070

1.786%

1.889%

3.403%

7/10

2.362

1.967%

2.296%

3.879%

8/10

2.693

3.310%

3.515%

4.859%

9/10

3.018

3.962%

5.175%

6.200%

10/10

3.350

9.051%

9.318%

10.309%